Modern designed treadmill specifically designed for children. It can be controlled by external ECG and pulmonary devices. The emergency stop magnet contact on the standard front handrail provides additional safety for the user. The Valiant 2 Pediatric offers a smooth acceleration from 0 km/h and is continuously adjustable in a range of 0,1 – 12 km/h (0.06 – 7.5 mph). The running surface is 50 x 127 cm and the standard elevation is 0 - 25%. This treadmill is supplied without any display.
A USB A-B cable for service purposes as well as connecting to ECG and pulmonary testing devices will be standard delivered with the product.
To connect older ECG and pulmonary testing devices with RS232 or other connectors you need a special interface cable that can be ordered separately.
Maximum speed | 12 km/h | 7.5 mph | maximum speed of the treadmill |
Minimum operational speed | 0.1 km/h | 0.1 mph | minimum speed of the treadmill |
Positive elevation | 25 % | maximum positive angle as a percentage | |
Elevation adjustment steps | 0.5 % | the size of the steps you can adjust the elevation |
Speed accuracy | 5 % | accuracy of speed | |
Accuracy inclination | 0,5 % | Accuracy of the inclination |
Allowed user weight | 225 kg | 496 lbs | maximum patient weight |
USB connector | connection via USB possible | ||
RS232 in connector | connection via RS232 |
Screen resolution | 800 x 400 pixels | Screen resolution | |
Walking surface length | 127 cm | 50 inch | length of the walking surface |
Walking surface width | 50 cm | 19.7 inch | width of the walking surface |
Step up height | 17 cm | 6.7 inch | height to step up the treadmill |
Product length (cm) | 192 cm | 75.6 inch | length of product in cm |
Product width (cm) | 76 cm | 29.9 inch | width of the product in cm |
Product height | 130 cm | 51.2 inch | height of product |
Power cord length | 250 cm | 98.4 inch | length of powercord including plugs |
Power cord IEC 60320 C19 with CEE 7/7 plug | European power cord and connector standard supplied with product | ||
Power cord NEMA | NEMA power cord and connector | ||
Maximum rated power input | 3000 VA | power input | |
115 V AC 50/60 Hz (2 phases) | when <230V, the treadmill should be connected between two phases | ||
230 V AC 50/60 Hz | operational power | ||
Maximum motor power | 1.8 kW | the maximum power the motor will consume |
IEC 60601-1:2005 | the product is IEC 60601-1 3rd edition compliant | ||
ISO 13485:2016 compliant | Lode fulfils ISO 13485:2016 requirements | ||
ISO 9001:2015 compliant | Lode fulfills ISO 9001: 2015 requirements | ||
Standard emergency lanyard | an emergency stop with a magnetic lanyard is present |
CE class Im according to MDD93/42/EEC | CE certified | ||
CE class of product with optional SpO2 | IIa | CE class of product with optional SpO2 | |
CE class of product with optional BPM | IIa | CE class of product with optional Blood Pressure Measurement | |
CB according to IECEE CB | CB certified |
Minimum operational temperature | 14 °C | minimum temperature at which the device will work within specification | |
Maximum operational temperature | 40 °C | maximum temperature at which the device will work within specification | |
Minimum operational air pressure | 70 kPa | minimum airpressure as from which the product will be operating within specification | |
Maximum operational air pressure | 106 kPa | maximum air pressure as from which the product will be operating within specification | |
Minimum operational non-condensing humidity | 30 % | minimum non-condensing humidity at which the product will operate within specification | |
Maximum operational non-condensing humidity | 90 % | maximum non-condensing humidity at which the product will operate within specification | |
Minimum storage & transport temperature | -25 °C | minimum temperature the product may be stored and transported at | |
Maximum storage & transport temperature | 70 °C | maximum temperature the product may be stored or transported at | |
Minimum air pressure storage & transport | 50 kPa | minimum air pressure the product may be stored or transported at | |
Maximum storage & transport air pressure | 106 kPa | maximum air pressure the product may be stored and transported at | |
Min. humidity storage & transport | 10 % | minimum non-condensing humidity that the product can be stored or transport at | |
Max. humidity storage & transport | 95 % | max. non-condensing humidity that the product can be stored or transported at |