In this system, the Angio is supplied with a multifunctional fixation set. With this set, the Angio ergometer is equipped with pedal shoes and can be mounted easily on all Imaging / Physiotherapy / Examination tables that have a standard instrument side rail of 25x10 mm. The adjustability range is 625 - 903 mm (width) and 20-120 mm (height). *
The Angio rehab is standard supplied with a 7" control unit with touch screen. Thanks to the built-in network module, the ergometer can be connected to the Lode Cardiac Rehab Manager Software, Lode Rehab Manager or the Lode Ergometry Manager.
A USB A-B cable for service purposes only will be standard delivered with the product. To connect LEM or LCRM you need a special interface cable that can be ordered under part number 930930.
Note: always check the width of the examination table on wich the ergometer is going to be installed. There is a fixaton set available for tables that are smaller (#967820)
Workload range fixed torque | 0,1 - 70 Nm | Workload range in fixed torque mode | |
Minimum load | 7 W | the minimum load the ergometer can provide | |
Maximum peak load | 1000 W | the maximum load the ergometer can provide for a short period of time | |
Minimum load increments | 1 W | the smallest steps with which the load can be added | |
Maximum continuous load | 750 W | the maximum power the ergometer can deliver continuous | |
Hyperbolic workload control | control of workload in hyperbolic way | ||
Linear workload control | control ergometer in a linear way | ||
Fixed torque workload control | control ergometer in a fixed torque way | ||
Maximum rpm independent constant load | 150 rpm | rpm as to which constant load can be applied | |
Minimum rpm independent constant load | 30 rpm | rpm as from which constant load can be applied | |
Optional heart rate controlled workload | control the device optionally through a set heartrate | ||
Electromagnetic "eddy current" braking system | system that very accurately applies a brake to the ergometer | ||
Dynamic calibration | continuous calibration during application of workload |
Workload accuracy from 7 to 100 W | 3 W | accuracy of the ergometer between 7 and 100 Watt | |
Workload accuracy from 100 to 500 W | 3 % | accuracy of the ergometer between 100 and 500 Watt | |
Workload accuracy from 500 to 1000 W | 5 % | accuracy of the ergometer between 500 and 1000 Watt |
Adjustability range height | 20 - 120 mm | adjustability range height | |
Adjustability range width | 625 - 903 mm | 24.6 kJ | adjustability range width |
Pedal shoes | premounted pedal shoes |
English user interface | User interface available in English language | ||
Chinese user interface | User interface available in Chinese language | ||
Croatian user interface | User interface available in croatian language | ||
Czech user interface | User interface available in Czech language | ||
Danish user interface | User interface available in Danishlanguage | ||
Dutch user interface | User interface available in Dutch language | ||
Finnish user interface | User interface available in Finnish language | ||
French user interface | User interface available in French language | ||
German user interface | User interface available in German language | ||
Greek user interface | User interface available in Greek language | ||
Hungarian user interface | User interface available in Hungarian language | ||
Italian user interface | User interface available in Italian language | ||
Japanese user interface | User interface available in language | ||
Korean user interface | User interface available in Korean language | ||
Latvian user interface | User interface available in Latvian language | ||
Lithuanian user interface | User interface available in Lithuanian language | ||
Norwegian user interface | User interface available in Norwegian language | ||
Polish user interface | User interface available in Polish language | ||
Portugese user interface | User interface available in Portugese language | ||
Romanian user interface | User interface available in romanian language | ||
Russian user interface | User interface available in Russian language | ||
Spanish user interface | User interface available in Spanish language | ||
Swedish user interface | User interface available in Swedish language | ||
Turkish user interface | User interface available in Turkish language | ||
Ukrainian user interface | User interface available in Ukrainian language | ||
Readout Distance | Readout the distance on control unit | ||
Readout RPM | Readout the actual rpm on control unit | ||
Readout Heartrate | Readout the actual heartarte on control unit | ||
Readout target HR | Readout target heartarte on control unit (only possible with HR option) | ||
Readout Energy | Readout the energy on control unit | ||
Readout Torque | Readout torque on control unit | ||
Readout Time | Readout time on control unit | ||
Readout Power | Readout the actual power on control unit | ||
Set Display | Choosing parameters to be shown on the display during the stress test | ||
Set Resistance | Possibility to define a specific resistance for each test subject | ||
Set P-Slope | Possibility to set the load regulation resistance | ||
Set Mode | Possibility to define parameters to be adjusted during the stress test | ||
Manual operation mode | Operate manually | ||
Preset protocol operation mode | Operate via preset protocols | ||
Analog operation mode | operate analog | ||
Terminal operation mode | operate with an external device | ||
External control unit | a control unit is connected with a cable to the ergometer | ||
Selfdesigned protocol operation mode | operate via selfdesigned protocols |
Maximum table width | 920 mm | 36.2 inch | maximum distance between the outside of the instrument rails on a patient table |
Minimum table width | 625 mm | 24.6 inch | minimum distance between the outside of the instrument rails on a patient table |
Product length (cm) | 54 cm | 21.3 inch | length of product in cm |
Product width (cm) | 68 cm | 26.8 inch | width of the product in cm |
Product height | 73 cm | 28.7 inch | height of product |
Product weight | 40 kg | 88.2 lbs | the weight of the product |
V AC | 100 - 240 V | voltage in Volt | |
Phases | 1 | phases | |
Frequency | 50/60 Hz | frequency in Herz | |
Power consumption | 160 W | power consumption in Watt | |
Power cord IEC 60320 C13 with CEE 7/7 plug | European power cord and connector standard supplied with product | ||
Power cord NEMA | NEMA power cord and connector |
IEC 60601-1:2012 | the product is IEC 60601-1 edition 3.1 compliant | ||
ISO 13485:2016 compliant | Lode fulfils ISO 13485:2016 requirements | ||
ISO 9001:2015 compliant | Lode fulfills ISO 9001: 2015 requirements |
CE class Im according to MDD93/42/EEC | CE certified | ||
CE class of product with optional SpO2 | IIa | CE class of product with optional SpO2 | |
CE class of product with optional BPM | IIa | CE class of product with optional Blood Pressure Measurement | |
CB according to IECEE CB | CB certified |
Minimum operational temperature | 14 °C | minimum temperature at which the device will work within specification | |
Maximum operational temperature | 35 °C | maximum temperature at which the device will work within specification | |
Minimum operational air pressure | 80 kPa | minimum airpressure as from which the product will be operating within specification | |
Maximum operational air pressure | 106 kPa | maximum air pressure as from which the product will be operating within specification | |
Minimum operational non-condensing humidity | 30 % | minimum non-condensing humidity at which the product will operate within specification | |
Maximum operational non-condensing humidity | 90 % | maximum non-condensing humidity at which the product will operate within specification | |
Minimum storage & transport temperature | -20 °C | minimum temperature the product may be stored and transported at | |
Maximum storage & transport temperature | 60 °C | maximum temperature the product may be stored or transported at | |
Minimum air pressure storage & transport | 50 kPa | minimum air pressure the product may be stored or transported at | |
Maximum storage & transport air pressure | 106 kPa | maximum air pressure the product may be stored and transported at | |
Min. humidity storage & transport | 10 % | minimum non-condensing humidity that the product can be stored or transport at | |
Max. humidity storage & transport | 95 % | max. non-condensing humidity that the product can be stored or transported at |