The Calibrator Ultra is an easy to set up and very accurate portable calibration station. The calibration station is controlled by a PC or laptop. With this station it is possible to control or calibrate all existing ergometers accurately. The Lode Calibrator is mounted on the axis of the ergometer. This system is approved by an independent institute, and is used by calibration institutes all over the world since 1993. Calibrations can be done up to 1500 watt. Due to the possibilities you create when using a normal PC to calibrate the calibrator you can print a calibration report for your customer and/or create your own database of calibration data.
The Calibrator Ultra is a professional Calibration station, made by Lode, which helps you to professionalize your calibrations and make your service more efficient.
Please note, specifications at 110 V are different than at 230 V.
Maximum rotational speed at 110 V | 120 rpm | Calibrator maximum rotations at 110 V | |
Minimum rotational speed | 25 rpm | calibrator minimum rotations | |
Maximum rotational speed | 180 rpm | calibrator maximum rotations | |
Rotation direction clockwise | looking at the outgoing axis | ||
Rotation direction anti-clockwise | looking at the outgoing axis | ||
Maximum motor torque | 100 Nm | the maximum torque the motor can deliver | |
Maximum sensor torque | 100 Nm | the maximum torque the sensor can measure | |
Warming up time | 5 min | the time to bring the product on operational temperature | |
Approved power range (maximum) | 1000 W | power range for which use is approved | |
Approved power range (minimum) | 0 W | power range for which use is approved | |
Power range at maximum rpm (maximum) | 1500 W | operational power range | |
Power range at maximum rpm (minimum) | 0 W | operational power range | |
Crank adapter included | to connect crank to calibrator |
Speed accuracy | 0.25 % | accuracy of speed | |
Power accuracy error > 100 W | 1 % | accuracy of power application | |
Power accuracy error up to 100W | 1 W | accuracy of power application | |
Torque resolution | 0.05 Nm | resolution of the torque adjustment | |
Time resolving | 10 ms | resolving time of the calibrator | |
Accuracy rpm adjustment | 0.1 rpm | the accuracy of the rotational speed adjustment | |
Rpm resolution | 0.05 rpm | resolution of the rotational speed |
Software available in English | Software available in English | ||
Software available in German | Software available in German | ||
Software available in Dutch | Software available in Dutch | ||
Software available in Japanese | Software available in Japanese |
USB connector | connection via USB possible |
Cable length | 3000 mm | 118.1 inch | length of cable |
Product length (cm) | 52 cm | 20.5 inch | length of product in cm |
Product width (cm) | 32 cm | 12.6 inch | width of the product in cm |
Product height | 52 cm | 20.5 inch | height of product |
Product weight | 29 kg | 63.9 lbs | the weight of the product |
230 VAC | 14 A | Power requirement at 230 VAC | |
110 VAC | 15 A | Power requirements at 110 VAC | |
Phases | 1 | phases | |
Frequency | 50/60 Hz | frequency in Herz |
Integrated emergency stop | emergency stop fully integrated |
Minimum operational temperature | 5 °C | minimum temperature at which the device will work within specification | |
Maximum operational temperature | 40 °C | maximum temperature at which the device will work within specification | |
Minimum operational air pressure | 80 kPa | minimum airpressure as from which the product will be operating within specification | |
Maximum operational air pressure | 106 kPa | maximum air pressure as from which the product will be operating within specification | |
Minimum operational non-condensing humidity | 30 % | minimum non-condensing humidity at which the product will operate within specification | |
Maximum operational non-condensing humidity | 90 % | maximum non-condensing humidity at which the product will operate within specification | |
Minimum storage & transport temperature | -20 °C | minimum temperature the product may be stored and transported at | |
Maximum storage & transport temperature | 60 °C | maximum temperature the product may be stored or transported at | |
Minimum air pressure storage & transport | 50 kPa | minimum air pressure the product may be stored or transported at | |
Maximum storage & transport air pressure | 106 kPa | maximum air pressure the product may be stored and transported at | |
Min. humidity storage & transport | 10 % | minimum non-condensing humidity that the product can be stored or transport at | |
Max. humidity storage & transport | 95 % | max. non-condensing humidity that the product can be stored or transported at | |
Mìnimum safe temperature | 5 °C | Minimum ambient temperature at which the device can be used safely | |
Maximum safe temperature | 40 °C | Maximum ambient temperature at which the device can be used safely |
Internal RAM | 4000 MB | minimum PC internal memory | |
Minimum Free Hard Disk space | 25 GB | software and storage size requirement | |
Number of free USB ports | 2 | for connection purposes |
Suitable for all brands | Calibrator can be used for ergometers of all brands | ||
Suitable for high power ergometers | Calibrator can be used for ergometers with a max. worload of >1000 W | ||
Microsoft Windows 10 | compatible with Microsoft Windows 10 |
PC software included | software needed for operation included |