In this system, the Angio is supplied with a multifunctional fixation set. With this set, the Angio ergometer is equipped with pedal shoes and can be mounted easily on all Imaging / Physiotherapy / Examination tables that have a standard instrument side rail of 25x10 mm. The adjustability range is 625 - 903 mm (width) and 20-120 mm (height). *
The Angio imaging is an ergometer that can be used for both sitting and supine ergometry. Its compact design makes it universally applicable for ergometric studies in those sectors in which standard ergometry cannot be used. The Angio operates independent of pedaling speed in the range of 7 - 1000 watt. The Angio imaging is standard supplied with a communication module and can therefor be easily controlled by all known stress ECG and pulmonary devices in the world. The workload, rpm and time can be readout from the 7" colour display.
* Note: always check the width of the examination table on which the ergometer is going to be installed. There is a fixaton set available for tables that are smaller (#967820).
The Angio imaging with fixation set is standard equipped with pedal shoes.
Workload range fixed torque | 0,1 - 70 Nm | Workload range in fixed torque mode | |
Minimum load | 7 W | the minimum load the ergometer can provide | |
Maximum peak load | 1000 W | the maximum load the ergometer can provide for a short period of time | |
Minimum load increments | 1 W | the smallest steps with which the load can be added | |
Maximum continuous load | 750 W | the maximum power the ergometer can deliver continuous | |
Hyperbolic workload control | control of workload in hyperbolic way | ||
Linear workload control | control ergometer in a linear way | ||
Fixed torque workload control | control ergometer in a fixed torque way | ||
Maximum rpm independent constant load | 150 rpm | rpm as to which constant load can be applied | |
Minimum rpm independent constant load | 30 rpm | rpm as from which constant load can be applied | |
Optional heart rate controlled workload | control the device optionally through a set heartrate | ||
Electromagnetic "eddy current" braking system | system that very accurately applies a brake to the ergometer | ||
Dynamic calibration | continuous calibration during application of workload |
Workload accuracy from 7 to 100 W | 3 W | accuracy of the ergometer between 7 and 100 Watt | |
Workload accuracy from 100 to 500 W | 3 % | accuracy of the ergometer between 100 and 500 Watt | |
Workload accuracy from 500 to 1000 W | 5 % | accuracy of the ergometer between 500 and 1000 Watt |
Shoe size pedal shoe EU | 32 - 41 | Shoe size belonging to pedal size EU | |
Shoe size pedal shoe US male | 1 - 8 | Shoe size belonging to pedal size US male | |
Shoe size pedal shoe US female | 1 - 9 | Shoe size belonging to pedal size US female | |
Shoe size pedal shoe UK | 12.5 (CH) - 7.5 (ad) | Shoe size belonging to pedal size UK | |
Adjustability range height | 20 - 120 mm | adjustability range height | |
Adjustability range width | 640 - 903 mm | 25.2 kJ | adjustability range width |
Pedal shoes | premounted pedal shoes |
English user interface | User interface available in English language | ||
Chinese user interface | User interface available in Chinese language | ||
Croatian user interface | User interface available in croatian language | ||
Czech user interface | User interface available in Czech language | ||
Danish user interface | User interface available in Danishlanguage | ||
Dutch user interface | User interface available in Dutch language | ||
Finnish user interface | User interface available in Finnish language | ||
French user interface | User interface available in French language | ||
German user interface | User interface available in German language | ||
Greek user interface | User interface available in Greek language | ||
Hungarian user interface | User interface available in Hungarian language | ||
Italian user interface | User interface available in Italian language | ||
Japanese user interface | User interface available in language | ||
Korean user interface | User interface available in Korean language | ||
Latvian user interface | User interface available in Latvian language | ||
Lithuanian user interface | User interface available in Lithuanian language | ||
Norwegian user interface | User interface available in Norwegian language | ||
Polish user interface | User interface available in Polish language | ||
Portugese user interface | User interface available in Portugese language | ||
Romanian user interface | User interface available in romanian language | ||
Russian user interface | User interface available in Russian language | ||
Spanish user interface | User interface available in Spanish language | ||
Swedish user interface | User interface available in Swedish language | ||
Turkish user interface | User interface available in Turkish language | ||
Ukrainian user interface | User interface available in Ukrainian language | ||
Readout Distance | Readout the distance on control unit | ||
Readout RPM | Readout the actual rpm on control unit | ||
Readout Heartrate | Readout the actual heartarte on control unit | ||
Readout target HR | Readout target heartarte on control unit (only possible with HR option) | ||
Readout Energy | Readout the energy on control unit | ||
Readout Torque | Readout torque on control unit | ||
Readout Time | Readout time on control unit | ||
Readout Power | Readout the actual power on control unit | ||
Set Display | Choosing parameters to be shown on the display during the stress test | ||
Set Resistance | Possibility to define a specific resistance for each test subject | ||
Set P-Slope | Possibility to set the load regulation resistance | ||
Set Mode | Possibility to define parameters to be adjusted during the stress test | ||
Manual operation mode | Operate manually | ||
Preset protocol operation mode | Operate via preset protocols | ||
Terminal operation mode | operate with an external device | ||
External control unit | a control unit is connected with a cable to the ergometer |
Maximum table width | 920 mm | 36.2 inch | maximum distance between the outside of the instrument rails on a patient table |
Minimum table width | 640 mm | 25.2 inch | minimum distance between the outside of the instrument rails on a patient table |
Product length (cm) | 54 cm | 21.3 inch | length of product in cm |
Product width (cm) | 68 cm | 26.8 inch | width of the product in cm |
Product height | 73 cm | 28.7 inch | height of product |
Product weight | 40 kg | 88.2 lbs | the weight of the product |
V AC | 100 - 240 V | voltage in Volt | |
Phases | 1 | phases | |
Frequency | 50/60 Hz | frequency in Herz | |
Power consumption | 160 W | power consumption in Watt | |
Power cord IEC 60320 C13 with CEE 7/7 plug | European power cord and connector standard supplied with product | ||
Power cord NEMA | NEMA power cord and connector |
IEC 60601-1:2012 | the product is IEC 60601-1 edition 3.1 compliant | ||
ISO 13485:2016 compliant | Lode fulfils ISO 13485:2016 requirements | ||
ISO 9001:2015 compliant | Lode fulfills ISO 9001: 2015 requirements |
CE class Im according to MDD93/42/EEC | CE certified | ||
CE class of product with optional SpO2 | IIa | CE class of product with optional SpO2 | |
CE class of product with optional BPM | IIa | CE class of product with optional Blood Pressure Measurement | |
CB according to IECEE CB | CB certified |
Minimum operational temperature | 14 °C | minimum temperature at which the device will work within specification | |
Maximum operational temperature | 35 °C | maximum temperature at which the device will work within specification | |
Minimum operational air pressure | 80 kPa | minimum airpressure as from which the product will be operating within specification | |
Maximum operational air pressure | 106 kPa | maximum air pressure as from which the product will be operating within specification | |
Minimum operational non-condensing humidity | 30 % | minimum non-condensing humidity at which the product will operate within specification | |
Maximum operational non-condensing humidity | 90 % | maximum non-condensing humidity at which the product will operate within specification | |
Minimum storage & transport temperature | -20 °C | minimum temperature the product may be stored and transported at | |
Maximum storage & transport temperature | 60 °C | maximum temperature the product may be stored or transported at | |
Minimum air pressure storage & transport | 50 kPa | minimum air pressure the product may be stored or transported at | |
Maximum storage & transport air pressure | 106 kPa | maximum air pressure the product may be stored and transported at | |
Min. humidity storage & transport | 10 % | minimum non-condensing humidity that the product can be stored or transport at | |
Max. humidity storage & transport | 95 % | max. non-condensing humidity that the product can be stored or transported at |